Chapter 3 Programming Your Universal Counter for Remote Operation
Command Settings for Optimizing Throughput
Set reference oscillator to non-auto state (internal or external):
[:SENSe]:ROSCillator:SOURce INTernal | EXTernal
Disable checking of external source if using external reference oscillator:
[:SENSe]:ROSCillator:EXTernal:CHECk OFF
Disable automatic interpolator calibration:
:DIAGnostic:CALibration:INTerpolator:AUTO OFF
Disable display:
:DISPlay:ENABle OFF
Disable printing:
:HCOPy:CONTinuous OFF
Disable post-processing (math, limit testing, statistics):
:CALCulate:MATH:STATe OFF
:CALCulate2:LIMit:STATe OFF
:CALCulate3:AVERage[:STATe] OFF
Specify expected frequency for Frequency, Period, Ratio, Phase, and Duty
Cycle measurements:
[:SENSe]:FREQuency:EXPected[1|2|3] <numeric_value> [HZ]
(See Note below.)
Specify ASCII format for result query responses:
:FORMat[:DATA] ASCii
Specify continuous measurements:
:INITiate:CONTinuous ON
Configure the read/fetch function memory:
Issue the following query and read the response.
:FETCh[:SCALar]:<function>?
__________________________
Note: This value is not stored on Save and is not part of the non-volatile state.

Typical Optimizing Throughput Results for Different Computers

3-36
(See Note below.)
(See Note below.)
(See Note below.)
(See Note below.)
Programming Guide