HP 53131A/132A 225 MHz Programming Manual page 174

225 mhz universal counter
Table of Contents

Advertisement

Chapter 4 Command Reference
:CALCulate[1] Subsystem
:CALCulate[1]:FEED "[:]SENSe[1]"
Sets or queries the data flow to be fed into the CALCulate[1] block.
Since the Counter can only sense one function at a time, there is only one valid
parameter.
Query Response
The string "SENS" is returned.
Comments
*RST: "SENSe[1]"
:CALCulate[1]:IMMediate
This command is an event that causes the Counter to recalculate existing data
without re-acquiring data. (This recalculation also happens automatically when
any change is made to the
:CALCulate[1|2] subsystems while :CALC:IMM:AUTO is ON.)
:CALC:IMM? is semantically equivalent to :CALC:IMM;DATA?.
The query form outputs the results of the new calculation.
This command will not affect:
Query Response
Comments
This command causes post-processing to occur in the :CALCulate2 subsystem, as
well as the CALCulate subsystem.
4-8
:CALC2:LIM:FCO
:CALC2:LIM:PCO
:CALC3: ...
Result will be formatted according to :FORMat[:DATA] ASCii | REAL
setting.
When ASCii format is used, numeric data is transferred as ASCII bytes in
<NR3> format. The number of significant digits will range from 1 to 15,
depending on the measurement resolution.
Only significant digits will be returned.
If no valid result exists, Not a Number 9.91E37 is returned and error -230 is
generated.
If the current measurement is Totalize or Voltage Peaks, Not a Number
9.91E37 is returned and error -221 is generated.
Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

53131a53132a

Table of Contents