Response Message Data Types - HP 53131A/132A 225 MHz Programming Manual

225 mhz universal counter
Table of Contents

Advertisement

Type
Description
<NR1>
This numeric representation has an implicit radix point.
The maximum number of characters in <NR1> response data is 17 (maximum
16 digits, 1 sign).
<NR2>
This numeric representation has an explicit radix point.
The maximum number of characters in <NR2> response data is 17 (maximum
15 mantissa digits, 1 sign, 1 decimal point).
<NR3>
This numeric representation has an explicit radix point and an exponent.
The maximum number of characters in <NR3> response data is 22 (maximum
15 mantissa digits, 2 signs, 1 decimal point , 1 ' E' character, 3 exponent digits).
Not a Number
Not a Number is represented by the value 9.91E37 . (Not a Number is defined in
IEEE 754). The Counter responds with this numeric value when queried for a
floating point number it cannot provide. This value will be formatted as an
<NR3>.
Chapter 3 Programming Your Universal Counter for Remote Operation
Elements of SCPI Commands

Response Message Data Types

Table 3-3 contains explanations of response data types.
Table 3-3. Response Message Data Types
Programming Guide
3-17