Chapter 4 Command Reference
[:SENSe] Subsystem
[:SENSe]:TOTalize:ARM:STOP:SOURce IMMediate |
EXTernal TIMer
Sets or queries the stop arm for Totalize measurements.
Query Response
A sequence of ASCII-encoded bytes: IMM, EXT or TIM
Comments
Related
Gate & ExtArm
Front-Panel
Keys
[:SENSe]:TOTalize:ARM:STOP:TIMer
<numeric_value> [S]
Sets or queries the gate time used in arming Totalize measurements.
<numeric_value>
Range
<numeric_value
Resolution
Query Response
Numeric data transferred as ASCII bytes in <NR3> format with six significant
digits.
Comments
Related
Gate & ExtArm
Front-Panel
Keys
4-106
*RST: TIMer
When both start and stop arm are set to IMMediate, use :INIT[:IMM] to
begin totalizing and :ABORt to terminate. No valid totalize result will exist
until the "auto" totalize measurement is terminated.
For short gate time: 1E-3 to 99.99E-3 seconds
For long gate time: 10 0E-3 to 1000.000 seconds
For short gate time: 0.01E-3 seconds
For long gate time: 1E-3 seconds
*RST: 100E-3 S
Only applies when [:SENS]:TOT:ARM:STOP:SOUR TIM is selected.