Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Gate & ExtArm Key to SCPI Command Map (Cont.)
Totalize
Auto Arming:
a.
[:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate
[:SENSe]:TOTalize:ARM:STOP:SOURce IMMediate
Time Arming:
b 1 . [:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate
b 2 . [:SENSe]:TOTalize:ARM:STOP:SOURce TIMer
c.
[:SENSe]:TOTalize:ARM:STOP:TIMer <numeric_value> [S]
External Arming:
d.
[:SENSe]:TOTalize:ARM[:STARt]:SOURce EXTernal
e.
[:SENSe]:TOTalize:ARM[:STARt]:SLOPe POSitive | NEGative
f 1 .
[:SENSe]:TOTalize:ARM:STOP:SOURce TIMer
f 2 .
[:SENSe]:TOTalize:ARM:STOP:SOURce EXTernal
[:SENSe]:TOTalize:ARM:STOP:SLOPe POSitive | NEGative
g.
[:SENSe]:TOTalize:ARM:STOP:TIMer <numeric_valu e> [S]
Rise Time, Fall Time, +/- Pulse Width, Dutycycle
Auto Arming:
a.
[:SENSe]:TINTerval:ARM[:STARt]:SOURce IMMediate
External Arming:
b.
[:SENSe]:TINTerval:ARM[:STARt]:SOURce EXTernal
c.
[:SENSe]:TINTerval:ARM[:STARt]:SLOPe POSitive | NEGative
Figure 2-4. Gate & ExtArm Key to SCPI Command Map (Part 3 of 6)