HP 53131A/132A 225 MHz Programming Manual page 271

225 mhz universal counter
Table of Contents

Advertisement

Chapter 4 Command Reference
Introduction
[:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate | EXTernal
Sets or queries the start arm for Totalize measurements.
Query Response
A sequence of ASCII-encoded bytes: IMM or EXT
Comments
Related
Gate & ExtArm
Front-Panel
Keys
[:SENSe]:TOTalize:ARM:STOP:SLOPe POSitive | NEGative
Sets or queries the slope of the external stop arm signal used in external arming
Totalize measurements.
Query Response
A sequence of ASCII-encoded bytes: POS or NEG
Comments
Related
Gate & ExtArm
Front-Panel
Keys
Programming Guide
*RST: IMMediate
When both start and stop arm are set to IMMediate, use :INIT[:IMM] to
begin totalizing and :ABORt to terminate. No valid totalize result will exist
until the "auto" totalize measurement is terminated.
*RST: NEGative
Only applies when [:SENS]:TOT:ARM:STOP:SOUR EXT is selected.
4-105

Advertisement

Table of Contents
loading

This manual is also suitable for:

53131a53132a

Table of Contents