Number
Error String
-400
Query error
-410
Query INTERRUPTED
-420
Query UNTERMINATED
-430
Query DEADLOCKED
-440
Query UNTERMINATED
after indefinite response
+2000
Offset calibration on channel 1 failed
+2001
Offset calibration on channel 2 failed
+2002
Gain calibration on channel 1 failed
+2003
Gain calibration on channel 2 failed
+2004
Interpolator calibration failed
+2005
Oscillator calibration failed
+2006
Time interval calibration failed
+2007
Measurement hardware calibration
failed
+2008
Measurement interpolator calibration
failed
+2009
No valid external timebase
+2010
External timebase failed during
measurement
+2011
Recall setup failed; hardware failure
Recall setup failed; empty register
+2012
Save setup failed
+2013
EEPROM failed
Chapter 5 Errors
Error Types
Table 5-2. Errors (Continued)
Cause
This is the generic query error.
Indicates that a condition causing an INTERRUPTED Query
error occurred. For example, a query followed by DAB or GET
before a response was completely sent.
Indicates that a condition causing an UNTERMINATED Query
error occurred. For example, the Counter was addressed to talk
and an incomplete program message was received.
Indicates that a condition causing a DEADLOCKED Query error
occurred. For example, both input buffer and output buffer are
full and the Counter cannot continue.
Indicates that a query was received in the same program
message after a query requesting an indefinite response
( for example, *IDN? or *OPT?) was executed.
:DIAGnostic:CALibration:INP:OFFS:AUTO ONCE failed.
:DIAGnostic:CALibration:INP2:OFFS:AUTO ONCE failed.
:DIAGnostic:CALibration:INP:GAIN:AUTO ONCE failed.
:DIAGnostic:CALibration:INP2:GAIN:AUTO ONCE failed.
:DIAGnostic:CALibration:INTerpolator:AUTO ONCE, *CAL?,
or :CALibration[:ALL]? failed.
:DIAGnostic:CALibration:ROSCillator:AUTO ONCE failed.
:DIAGnostic:CALibration:TINTerval:QUICk failed.
A measurement calibration failed on the last measurement.
Interpolator calibration failed on the last measurement;
therefore, no valid measurement was taken.
You have selected external reference and there is no external
reference applied to the rear-panel Ref In connector, or the
external signal is not an allowed frequency.
You have selected the auto reference mode and the Counter
detected that the external reference became invalid
measurement. Therefore, the current result is not valid, and the
Counter switches to using the internal reference.
*RCL failed.
*RCL failed because the register specified for recall is empty.
*SAV failed, or *RCL couldn ' t save to register 0.
A request to update a setting which is stored in the EEPROM
(:SYST:COMM:SER:TRAN:BAUD,
:SYST:COMM:SER:TRAN:PAR,
:SYST:COMM:SER:TRAN:PACE,
:SYST:COMM:SER:CONT:DTR, :DISP[:WIND]:TEXT:RAD,
:CAL[:DATA], or :DIAG:CAL: ...) resulted in a hardware failure.