HP 53131A/132A 225 MHz Programming Manual page 57

225 mhz universal counter
Table of Contents

Advertisement

Keyword/Syntax
[:SENSe] (Cont.)
:EVENt3
:LEVel
[:ABSolute]?
:SLOPe?
:FREQuency
:ARM
[:STARt]
:SLOPe
:SOURce
:STOP
:DIGits
:SLOPe
:SOURce
:TIMer
:EXPected[1|2|3]
:AUTO
Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Parameter Form
POSitive | NEGative
IMMediate | EXTernal
<numeric_value>
POSitive | NEGative
IMMediate | EXTernal
|
TIMer | DIGits
<numeric_value> [S]
<numeric_value> [HZ]
ON
Programming Guide
Std/
Comments
New
New
Subtree. Queries the characteristics of the " trigger
event" for channel 3 input.
New
Subtree.
New
Query only. Returns the channel 3 input trigger level.
New
Query only. Returns the edge of the channel 3 input
that will be considered an event.
Std
Subtree. Controls the frequency, frequency ratio, and
period measuring capabilities of the instrument.
New
Subtree. Synchronizes the frequency start and stop
arm with events.
New
Subtree.
New
Sets the slope of the external start arm signal used in
external arming frequency, frequency ratio, and
period measurements. Only applies when
[:SENS]:FREQ:ARM[:STAR]:SOUR EXT is selected.
New
Sets the start arm for frequency, frequency ratio, and
period measurements.
New
Subtree.
New
Sets the resolution in terms of digits used in arming
frequency, frequency ratio, and period
measurements. Only applies when
[:SENS]:FREQ:ARM:STOP:SOUR DIG is selected.
New
Sets the slope of the external stop arm signal used in
external arming frequency, frequency ratio, and
period measurements. Only applies when
[:SENS]:FREQ:ARM:STOP:SOUR EXT is selected.
New
Sets the stop arm for frequency, frequency ratio, and
period measurements.
New
Sets the gate time used in arming frequency,
frequency ratio, and period measurements. Only
applies when [:SENS]:FREQ:ARM:STOP:SOUR TIM
is selected.
New
Specifies the approximate frequency of a signal you
expect to measure at channel 1, 2, or 3.
New
Configures Counter to perform a pre-measurement
step to automatically determine the approximate
frequency of the measurement signal(s).
2-33

Advertisement

Table of Contents
loading

This manual is also suitable for:

53131a53132a

Table of Contents