How To Program The Counter To Display Results; Configuring The Counter' S Display; Commands For Displaying Non-Scaled/Offset Results - HP 53131A/132A 225 MHz Programming Manual

225 mhz universal counter
Table of Contents

Advertisement

Chapter 3 Programming Your Universal Counter for Remote Operation
Elements of SCPI Commands

How to Program the Counter to Display Results

Configuring the Counter ' s Display
The Counter has five different display modes:
1. Non-scaled/offset results — frequency, period, time interval, etc. This
display mode is used on power-up.
2. Scaled/offset results— results modified by scale and offset values
3. Limit graph — a graphical look that shows if a measurement is within
limits
4. Statistics— mean, min, max or standard deviation
5. Display Enable— All segments and LEDs (except Remote and SRQ) off
or on
The following command groupings show how to program the Counter to any of
the above display modes.

Commands for Displaying Non-Scaled/Offset Results

The following lines will always show the raw (non-scaled/offset) measurement
result.
:DISP:MENU OFF
:DISP:TEXT:FEED ' CALC2'
:CALC2:LIM:DISP NUMBER
:CALC:MATH:STATE OFF
:CALC:IMM
Note that :CALC2:LIM:DISP NUMBER will only show the raw result if the
command :DISP:TEXT:FEED ' CALC2' is also issued. These commands must be
issued in pairs.
Programming Guide
Clear any menu items that may be on
display.
Show the non-statistical result.
Use the numeric display mode.
Disable math so scale and offset not used.
Cause a calculation to be made to update display.
3-43