HP 53131A/132A 225 MHz Programming Manual page 182

225 mhz universal counter
Table of Contents

Advertisement

Chapter 4 Command Reference
:CALCulate2 Subsystem
Comments
Query only.
:CALCulate2:LIMit:LOWer[:DATA] <numeric_value>
[HZ | S |DEG]
Sets or queries the lower limit used for limit testing.
When the result is less than the lower limit, a fail is reported;
when the result is equal to the lower limit, a fail is not reported.
If math is enabled (:CALC:MATH:STATe ON), the limit value specified should
take into account that the limit testing is on measurements that have been scaled
and offset.
<numeric_value>
-9.9999990000E+12 to -1.0000000000E-13, 0.0000000000, +1.0000000000E-
Range
13 to +9.9999990000E+12.
<numeric_value>
11 digits
Resolution
Query Response
Numeric data transferred as ASCII bytes in <NR3> format with eleven significant
digits.
Comments
Related
Uppr & Lower
Front-Panel
Keys
4-16
*RST: 0.0000000000
This command couples :CALC3:LFIL:LOW to the same value.
Updating the lower limit value causes the limit counts (:CALC2:LIM:FCO,
:CALC2:LIM:PCO) to be cleared.
The front panel menu item is not always able to display all of the significant
digits of this value. When this is the case, the displayed value is different from
the actual value in that the displayed value has been rounded. However, using
the front panel Enter key, while this value is in the 11-digit display, will
update the actual value to the displayed (rounded) value.