Chapter 4 Command Reference
Introduction
How to Use the Measurement Instruction Commands (Cont.)
Use
:SENS:EVEN:LEV 0V
to set the trigger level to 0 Volts.
Use
:INITIATE
to start the measurement.
Use
:FETCH?
to query for result.

Firmware Revision Work-Around Commands

The following applies to HP 53131A instruments with Firmware Revision 3413:
The three commands listed below should NOT be used.
:MEASure[:SCALar][:VOLTage]:MAXimum? <parameters>[<source_list>]
:MEASure[:SCALar][:VOLTage]:MINimum? <parameters>[<source_list>]
:MEASure[:SCALar][:VOLTage]:PTPeak? <parameters>[<source_list>]
The work-around commands which should be substituted are listed below.
:CONFigure[:SCALar][:VOLTage]:MAXimum <parameters>[<source_list>]
:INIT; *WAI; FETCh?
:CONFigure[:SCALar][:VOLTage]:MINimum <parameters>[<source_list>]
:INIT; *WAI; FETCh?
:CONFigure[:SCALar][:VOLTage]:PTPeak <parameters>[<source_list>]
:INIT; *WAI; FETCh?
The three commands listed below should NOT be used.
:READ[:SCALar][:VOLTage]:MAXimum?
:READ[:SCALar][:VOLTage]:MINimum?
:READ[:SCALar][:VOLTage]:PTPeak?
Programming Guide
4-77