Device Clear - HP 53131A/132A 225 MHz Programming Manual

225 mhz universal counter
Table of Contents

Advertisement

Chapter 4 Command Reference
Introduction

Device Clear

Device Clear
The full capability of the Device Clear IEEE 488.1 interface function is
implemented in the Counter. This function allows a device to be initialized to a
cleared state. The device-dependent effect is described below.
In response to either the Device Clear message or the Selected Device Clear
message, the Counter:
clears the input buffer and Output Queue,
resets the parser, execution control, and response formatter,
clears any command that would prevent processing a *RST or other
commands,
disables the effect of a prior *OPC command, and
terminates the holdoff action of a *WAI, *OPC?, or data query (:MEASure
query, :READ query, :FETCh query, :CALC:DATA?, :CALC2:LIM:FAIL?)
waiting for pending operation to complete.
Also, a front-panel initiated diagnostic or calibration may be aborted (for example,
if the front-panel diagnostic or calibration is waiting for user input).
Programming Guide
4-31