HP 53131A/132A 225 MHz Programming Manual page 220

225 mhz universal counter
Table of Contents

Advertisement

Chapter 4 Command Reference
Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)
:CONFigure?
Queries the function configured by the last :CONFigure or :MEASure query.
If the instrument state has changed through commands other than :CONFigure or
:MEASure query, the instrument will not track these changes, and the query
response will not reflect these changes.
Query Response
Comments
4-54
This command disables math, statistics, and limit-testing.
If an absolute trigger level is not specified in the <parameters>, then when this
command executes, for functions other than Voltage Peaks (maximum,
minimum, peak-to-peak) or Totalize,
– auto-trigger is enabled,
– auto-trigger level(s) are set,
– auto-trigger is invoked on measurement channel(s).
A string of the form: "<function> <parameters>[,<source_list>] ", omitting
the leading colon from the <function>.
The Rise Time, Fall Time, and Duty Cycle functions each have two ways of
being specified. This query response will use RTIM, FTIM, or DCYC to
indicate these functions.
The response is unaffected by *RST, recall, and [:SENS]:FUNC.
At power-on, this query generates an error and returns an empty string.
Refer to the sub-section in this section titled "Descriptions of Measurement
Functions " for descriptions of each measurement function.
Refer to Table 4-1 in this section for a summary of the <function>,
<parameters>, and <source_list> for each of the measurement functions.
Programming Guide