Chapter 4 Command Reference
Introduction
Descriptions of the Measurement Functions— <function> (Cont.)
:MEASure[:SCALar][:VOLTage]:FALL:TIME? (Cont.)
<source_list>
range:
default:
:MEASure[:SCALar][:VOLTage]:FREQuency?
[<expected_value>[,<resolution>]][, (@1)|(@2)|(@3)]
Measures Frequency.
The measurement arming mode is set to "digits." The Counter uses the
<expected_value> and <resolution> parameters to configure the number of digits
of resolution arming setting.
The Channel 1 or Channel 2 trigger settings are coupled so that the measurement
channel has auto-trigger enabled at 50% with a positive slope.
Ch1, Ch2
<expected_value>
range:
resolution:
default:
Ch1, Ch2 <resolution>
description:
range:
resolution:
Programming Guide
(@1)
(@1)
.100 Hz to 225 MHz
<expected value> should be within 10% of input
frequency for optimum arming configuration
10 MHz
value indicates decade corresponding to least significant
digit of the result
1E-16 to 1E6 Hz
value which indicates 3 to 15 digits of resolution for the
specified <expected value>
<resolution> should use a mantissa of 1.0 and be an
even power of 10