3-10. AC Current

Use a Fluke 5790A or equivalent with the appropriate precision shunts and adapter to
measure the 5500A output. Refer to the 5790A Operator Manual for operating
instructions and connections. Enter into the 5500A each of the measured values listed in
Table 3-5 when prompted to do so.
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19

3-11. AUX DC Volts

Measure the AUX output using a precision DMM. Enter into the 5500A the measured
values of each step listed in Table 3-6 when prompted to do so.
Step
1
2
3-8
Table 3-5. AC Current Calibration Steps
5500A Output
(AUX)
3.2999 mA
0.330 mA
3 mA
3 mA
300 µA
300 µA
300 µA
30 mA
30 mA
30 mA
300 mA
300 mA
300 mA
2 A
2 A
2 A
10 A
10 A
10 A
Table 3-6. AUX DC Volts Calibration Steps
NORMAL Output
+300 mV
+3 V
Frequency
100 Hz
100 Hz
5 kHz
10 kHz
100 Hz
5 kHz
10 kHz
100 Hz
5 kHz
10 kHz
100 Hz
5 kHz
10 kHz
100 Hz
1000 Hz
5 kHz
100 Hz
500 Hz
1000 Hz
Shunt Value
A40-10mA
1 kΩ Metal Film
A40-10mA
A40-10mA
1 kΩ Metal Film
1 kΩ Metal Film
1 kΩ Metal Film
A40-30mA
A40-30mA
A40-30mA
A40-300mA
A40-300mA
A40-300mA
A40-3A
A40-3A
A40-3A
Y5020, 0.01 Ω
Y5020, 0.01 Ω
Y5020, 0.01 Ω
AUX output
+300 mV
+3 V