1535/1537/1537-II
Users Manual
10 kV Clamp (Red/Black/Green)
Table 10. Number of Insulation Resistance Measurements
Un (V)
250 V
500 V
1000 V
2500 V
General Specifications
For a list of specifications for 1535/1537, please refer to the 1535/1537 Insulation
Tester Safety Information.
Electrical Specifications
The Tester's accuracy is specified for 1 year after calibration at operating
temperatures of 10 °C to 30 °C. For operating temperatures outside the range
(-10 °C to +10 °C and +30 °C to +50 °C), add ±0.25 % per °C for 5 % bands and
add ±1 % per °C for 20 % bands.
Test Voltage
250 V
28
Table 9. Accessories
Accessory
250 kΩ
500 kΩ
1 MΩ
2.5 MΩ
Table 11. Insulation Resistance Measurement
Range
<200 kΩ
200 kΩ to 500 kΩ
0.50 MΩ to 5.00 MΩ
5.0 MΩ to 50.0 MΩ
50 MΩ to 500 MΩ
0.50 GΩ to 5.00 GΩ
5.0 GΩ to 50.0 GΩ
>50 GΩ
R
(Ω)
load
Resolution
unspecified
1 kΩ
0.01 MΩ
0.1 MΩ
1 MΩ
0.01 GΩ
0.1 GΩ
unspecified
Part No.
4103525
Number of
Measurements
6500
3800
2200
1300
Accuracy
unspecified
5 %
5 %
5 %
5 %
5 %
20 %
unspecified