Interpolation: Linear, (sinx)/x

Triggering System

Modes: Normal, Auto, Single, and Stop
Sources: Any input channel, External, Ext/10, or line; slope and level are unique to each source (except line)
Coupling Mode: GND, DC 50 ohms, DC 1 Mohms, AC 1 Mohms
Pre-trigger Delay: 0 to 100% of memory size (adjustable in 1% increments or 100 ns)
Post-trigger Delay: 10,000 divisions in real time mode; limited at slower time/div settings
Holdoff by Time or Events: 1 ns to 20 s or from 1 to 99,999,999 events
Internal Trigger Range: ±4.1 div from center (typical)
Trigger and Interpolator Jitter: </= 3 ps
Maximum Trigger Sensitivity with Edge Trigger (Ch1-4 + external):
44Xi
2 div @ < 400 MHz
1 div @ < 200 MHz
Maximum Trigger Frequency with SMART Trigger (Ch1-4 + external):
44Xi
400 MHz
@ >/= 10 mV
Trigger Level DC Accuracy: ±4% of full scale ±2 mV (typical)
External Trigger Range: EXT/10 ±4 V; EXT ±400 mV

Basic Triggers

Edge/Slope/Line: Triggers when the signal meets the slope and level condition.
Width: Triggers on positive or negative pulse widths selectable from 500 ps to 20 s or on intermittent faults
(subject to bandwidth limit of oscilloscope).
Pattern: Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input – 2
Ch+EXT on WaveRunner 62Xi). Each source can be high, low, or don't care. The High and Low level can be
selected independently. Triggers at start or end of the pattern.
State or Edge Qualified: Triggers on any input source only if a defined state or edge occurred on another
input source. Delay between sources is selectable by time or events.
TV: Provides stable triggering on standard or custom composite video signals. Use them on PAL, SECAM, or
NTSC systems. Optional HDTV Trigger for 1080i, 1080p and 720p formats along with non-standard formats
up to 2000 lines.

SMART Triggers

Dropout: Triggers if the input signal drops out for longer than a selectable time-out between 1 ns and 20 s.
Glitch: Triggers on positive or negative glitches with widths selectable from 500 ps to 20 s or on intermittent
faults (subject to bandwidth limit of oscilloscope).
Signal or Pattern Interval: Triggers on intervals selectable from 1 ns to 20 s.
Runt: Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1
ns and 20 s.
Slew Rate: Activates a trigger when the rising or falling edge of a pulse crosses two threshold levels, an
WRXi-OM-E Rev C
rms
64Xi
2 div @ < 600 MHz
2 div @ < 600 MHz
1 div @ < 200 MHz
1 div @ < 200 MHz
64Xi
600 MHz
600 MHz
@ >/= 10 mV
@ >/= 10 mV
(typical)
62Xi
2 div @ < 1 GHz
1 div @ < 200 MHz
62Xi
1 GHz
@ >/= 10 mV
O
PERATOR
104MXi/Xi
2 div @ < 2 GHz
1 div @ < 200 MHz
104MXi/Xi
2 GHz
@ >/= 10 mV
'
M
S
ANUAL
204Xi
204Xi
17