Setting Up Help M
arkers
1. In the menu bar, touch Measure Setup...
2. Select a Measure Mode: Std Vertical, Std Horizontal, or My Measure.
3. Touch the Show All button to display Help Markers for every parameter being measured on the displayed
waveform (C2 in the examples above).
. Touch inside the Help Marker
4
Gate gate posts. The gate posts are independently placeable for each parameter. or Detailed The Detailed
selection produces cursors, Measure Gate gate posts, a label identifying the parameter being measured, and
a level indicator and hysteresis band for at level parameters.
Note: The choice of Simple or Detailed is applied to all parameters at the same time. That is, if you choose
Simple markers for one parameter, all parameters will be displayed in this mode.
5. Touch the Always On checkbox if you want to continuously display Help Markers for this parameter.

Turning Off Help Markers

1. Touch the Clear All button to turn off Help Markers for all parameters.
2. To turn off Help Markers for individual parameters, touch the Px tab for the parameter in question. Then
uncheck the Always On checkbox. When you close this dialog, the Help Markers for this parameter will no
longer be displayed.

Customizing a Parameter

From the Measure Dialog

1. Touch the My Measure button in the Measure dialog. The dialog presents you with a panel of eight preset
parameters.
2. For each parameter,
3. If you want to change the parameter listed, or a me
for example) alongside the check box. A pop-up menu of parameters categorized by type appears. To di
parameter icons only, touch the icon button
form, along with an explanation of each parameter, touch the list button
to scroll through the list of icons.
4. When you make a selection from the parameter icon menu, the setup dialogs for that parameter appear. You
can then change the waveform source and other conditions of the parameter.
5. If you are setting up an "@level" parameter, make selections for Level type (percent or absolute), Slope
(positive, negative, both), and Hysteresis level.
6. Touch the Gate tab, and set the position of the gate posts.
108
ERIES
s field and select Simple The Simple selection produces cursors and Measure
touch the On check box to enable the parameter listed.
asurement characteristic, touch the parameter button (P1
at the bottom of the menu. To display the icons in list
splay
. Use the Up/Down buttons
WRXi-OM-E Rev C