Statistics; Applying A Measure Mode; Measure Modes - LeCroy WaveRunner Xi series Operator's Manual

Oscilloscopes
Hide thumbs Also See for WaveRunner Xi series:
Table of Contents

Advertisement

W
R
X
S
AVE
UNNER
I
4. Select Measure
Parameter in error (P1) Out Result
5. Read the status information in line StatusDescription.

Statistics

By touching the Statistics On checkbox in the Measure, you can display statistics for standard vertical or
horizontal parameters, or for custom parameters. The statistics that are displayed are as follows:
The values displayed in the num row are the number of measurements computed. For
computes on an entire waveform (like edge@level, mean, minimum, maximum, etc.) th
represents the number of sweeps.
For any parameter that computes on every event
acquired waveform. If x waveforms were acquired, the value represents x times the number of cycles per
waveform. Also, the "value" is equal to the measurement of the last cycle on the last acquisition.

Applying a Measure Mode

1. In the menu bar, touch Measure, then Measure Setup.
2. Choose a Measure Mode from the dialog. The parameters are displayed below the grid.

Measure Modes

The selections for Measure Mode allow you to quickly apply parameters for standard vertical and standard
horizontal setups, and custom setups.
102
ERIES
value (last)
mean
min.
max.
sdev
num
, the value displayed is equal to the number of events per
any parameter that
e value displayed