LeCroy WaveRunner Xi series Operator's Manual page 114

Oscilloscopes
Hide thumbs Also See for WaveRunner Xi series:
Table of Contents

Advertisement

W
R
X
S
AVE
UNNER
I
Parameter
Description
Edge@level
Number of edges in waveform.
Excel
Performs measurements in Excel
by transferring one or two
waveforms and reading the
resulting parame
Fall time
Fall time: Duration of falling edge
from 90-10%.
Threshold arguments specify two
vertical values on each edge used
to compute fall time. Formulas for
upper and lower values:
lower = lower thresh. x amp/100 +
base
upper = upper thresh. x amp/100 +
base
Fall 80-20%
Fall 80-20%: Duration of pulse
waveform's falling transition from
80% to 20%, averag
transitions between the cursors.
Fall@level
Fall at level: Duration of pulse
waveform's falling edges betwe
user-specified transition levels. See
also Rise@level.
First
Indicates value of horizontal ax
left cursor.
114
ERIES
ter value.
ed for all falling
Definition
T
ime at upper threshold
m
inus Time at lower
th
reshold averaged
over each falling edge
A
verage duration of
falling
80-20% transition
D
uration of falling edge
en
between transition
le
vels
is at H
orizontal axis va
lue at
le
ft cursor
Notes
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate level
noise in data.
Available with JTA2 a
nd XMAP
options.
Available with XMAP option.
Excel must be loaded on the
instrument.
On signals not having two
levels (triangle or saw-tooth
waves, for example), top and
o maximum
base can default t
and minimum, giving,
however,
less predictable results.
Standard parameter.
On signals not having two
levels (triangle or saw-tooth
waves, for example), top and
base can d
efault to maximum
and minimum, giving,
however,
less predictable resul
ts.
Standard parameter.
On signals not having two ma
levels (
triangle or saw-tooth
waves, for example), top a
base can default to maximu
and minimum, giving, however,
less predictable results.
Standard parameter.
Indicates location of l
eft cursor.
Cursors are interchange
example, the left cursor may be
moved to the right of the right
cursor and first will give the
location of the cursor formerly on
the right, now on left.
Standard parameter.
WRXi-OM-E Rev C
s from
major
major
jor
nd
m
able: for

Advertisement

Table of Contents
loading