Show Last Trace

For most applications, you may not wa
persistence display. In those cases turn
doing mask testing and want to see wh

Persistence Time

You can control the duration of persistence by setting a time limit, in seconds, after which persistence data will be
erased: 0.5 s, 1 s, 2 s, 5 s, 10 s, 20 s, or infinity.

Persistence Setup

1. In the menu bar touch Display, then touch Persistence Setup... in the drop-down menu.
2. Touch the Persistence On checkbox.
3. Touch one of the mode buttons
4. Then touch the Show last trace checkbox if you want the last trace displayed.
5. Touch inside the Saturation field and enter a whole number integer, using the pop-up numeric keypad.
6. Touch inside the Persistence Time field and make a selection from the pop-up menu.
7. If you selected Per Trace, for each input channel touch its tab, then make selections of mode, saturation
level, persistence time, and last trace display in the same way as for All Locked.
To create a 3-dimensional view, touch the 3d button
1. Touch inside the Saturation field and enter a whole number integer, using the pop-up numeric keypad.
2. Touch inside the Persistence Time field and make a selection from the pop-up menu.
3. Under "3D settings," touch inside the Quality field and select an image quality from the pop-up menu: wire
frame, solid, or shaded.
4. For each axis, touch inside the field and enter a value from -90° to +90°.
To turn off persistence for all channels, uncheck the Persistence On checkbox.

Screen Saver

The Windows screen saver is activated
1. Minimize the instrument display by touching File in the menu bar, then Minimize in the drop-down menu.
2. Touch Start down in the task bar.
3. Touch Settings in the pop-up menu.
4. Touch Control Panel.
5. Touch Display.
6. Touch the Screen Saver tab.
WRXi-OM-E Rev C
nt to show the last trace because it will be superimposed on top of your
off Show Last Trace by touching the checkbox. However, if you are
ere the last trace is falling, turn
in the same way as for any PC.
O
PERATOR
Show Last Trace on.
. Then
'
M
S
ANUAL
75

Advertisement

Table of Contents
loading

Table of Contents