Setting Up Pass/Fail Testing

Initial Setup

1. Touch Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu.
2. Touch the Actions tab.
3. Touch the Enable Actions checkbox. This will cause t
waveform's passing or failing a test.
4. Touch the Su
mmary View to enable a line of text that shows concisely the status of your last waveform and
keeps a runn
ing count of how many sweeps have passed.
5. Touch inside the Pass If field, and select a Boolean condition from the pop-up menu.
If you want to set up the test to end after a finite number of sweeps, touch the Stop Test checkbox. Then
touch inside the After field and enter a value, using the pop-up numeric keypad.
Under "If", touch either the Pass
waveform's passing or failing the test.
Under "Then", touch the actions you want to occur: stop test, sound alarm, print result, emit pulse, or save the
waveform. If you want to have the results printed and your oscilloscope is not equipped with a printer, be sure
that the it is connected to a local or network printer. See Printing.
If you want to save your waveform automatically, touch the Save Setup. This will take you out of the current
dialog and will open the Save Waveform dialog. See Saving and Recalling Waveforms.
Test your Pass/Fail conditions by touching the Force Actions Once button. Press the Clear All button to quickly
uncheck all checkboxes if you want to change your selections.

Comparing a Single Parameter

1. Touch Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu.
2. Touch a Qx tab; a setup dialog for that position will open.
3. Touch inside the Source1 field and select a so ce from the pop-up menu.
4. Touch inside the Condition field in the main dialog and select ParamCompare:
5. Touch inside the Compare Values field and select All or Any from the pop-up menu:
.
By selecting All, the test will be true only if every waveform falls within the limit that you will set. By selecting
Any, the test will be true if just one waveform falls within the limit.