Download Print this page

Endress+Hauser Levelflex FMP56 Operating Instructions Manual page 73

Hart guided-wave radar
Hide thumbs Also See for Levelflex FMP56:

Advertisement

Levelflex FMP56, FMP57 HART
Endress+Hauser
Error
Echo lost diagnostic
message appears after the
supply voltage is switched on.
Device displays a level when
the tank is empty.
Wrong slope of the level over
the entire measuring range
13.2
Diagnostic information on local display
13.2.1
Diagnostic message
Faults detected by the self-monitoring system of the measuring instrument are displayed
as a diagnostic message in alternation with the measured value display.
Measured value display in alarm condition
X X X X X X X
X
X
20.50
X
1
Status signal
2
Status symbol (symbol for event level)
3
Status symbol with diagnostic event
4
Event text
5
Operating elements
Status signals
"Failure (F)" option
A device error has occurred. The measured value is no longer valid.
A0032902
"Function check (C)" option
The device is in the service mode (e.g. during a simulation).
A0032903
"Out of specification (S)" option
The device is operated:
S
• Outside its technical specifications (e.g. during startup or cleaning)
• Outside the configuration performed by the user (e.g. level outside the configured
A0032904
range)
"Maintenance required (M)" option
Maintenance is required. The measured value is still valid.
A0032905
Possible cause
Error in the echo tracking
Noise level too high during
initialization phase.
Incorrect probe length
Wrong bin property selected.
2
1
S
X X X X X X X
S801
Supply voltage
XX
Diagnostics and troubleshooting
Solution
Deactivate echo tracking (Expert
→ Sensor → Echo tracking → Evaluation
mode = History off).
Enter a new value for the Empty
calibration parameter (→  114).
• Perform a probe length correction
(Confirm probe length parameter
(→  143)).
• Carry out mapping over the entire
probe length when the tank is empty
(Confirm distance parameter
(→  117)).
Select the correct Bin type parameter
(→  113).
Diagnostic message
X
X
S
i
Menu
3
4
5
A0029426-EN
73

Advertisement

loading
Need help?

Need help?

Do you have a question about the Levelflex FMP56 and is the answer not in the manual?

This manual is also suitable for:

Levelflex fmp57