Operating menu
166
• Of customer settings
All customer parameters are reset to their factory setting. Service parameters, however,
remain unchanged.
• To transducer defaults
Every measurment-related parameter is reset to its factory setting. Service parameters
and communication-related parameters, however, remain unchanged.
• Restart device
The restart resets every parameter which is stored in the volatile memory (RAM) to the
factory setting (e.g. measured value data). The device configuration remains unchanged.
Levelflex FMP56, FMP57 HART
Endress+Hauser