12.6
Configuring the current outputs
12.6.1
Factor setting of current outputs for level measurements
Current output
Assigned measured
value
1
Level linearized
2
Relative echo
(For devices with two
amplitude
current outputs)
12.6.2
Adjusting the current outputs
The current outputs can be adjusted in the following submenus:
Basic setup
Setup → Advanced setup → Current output 1 to 2
Advanced settings
Expert → Output 1 to 2 → Current output 1 to 2
See "Description of Device Parameters" GP01000F
12.7
Configuration management
After commissioning, you can save the current device configuration, copy it to another
measuring point or restore the previous device configuration. You can do so using the
Configuration management parameter and the options available.
Path in the menu
Setup → Advanced setup → Configuration backup display → Configuration management
Meaning of the options
• Cancel
No action is executed and the user exits the parameter.
• Execute backup
A backup copy of the current device configuration is saved from the HistoROM
(integrated in the device) to the display module of the device.
• Restore
The last backup copy of the device configuration is copied from the display module to the
HistoROM of the device.
Commissioning via operating menu
4mA value
20mA value
0% or the corresponding
100% or the corresponding
linearized value
linearized value
0 mV
2 000 mV
69