Clearance Requirements; Electrical Requirements; Service Line Requirements - Labconco Cell Logic+ B2 User Manual

Biosafety cabinets
Hide thumbs Also See for Cell Logic+ B2:
Table of Contents

Advertisement

Clearance Requirements

A minimum clearance of at least 6 inches (150 mm) is suggested on the top and both
sides of the product for service.
See
Appendix B: Dimensions

Electrical Requirements

The product models have the following electrical requirements.
Catalog Number
3x349xx01
3x349xx21
3x349xx-11, 31, 41, 51, 61, 71
3x369xx01
3x369xx21
3x369xx-11, 31, 41, 51, 61, 71
1 Electrical Requirements, 'V' = VAC (Voltage with alternating current), 'A' = Amperes
A dedicated outlet with an appropriate circuit breaker should be located as close as
possible to the product, but no greater than 10 feet (3 m). Consult your local electrical
codes for properly rated circuit breakers. For safe operation the dedicated outlet must
provide a protective earthing ground connection to the product.
On 100V and 115V models, both internal electrical outlets are protected by a ground
fault interrupter circuit (GFIC). Labconco does NOT recommend connecting the
product's power cord into a GFIC outlet. GFIC outlets can nuisance trip, resulting in
complete removal of power to the product. Such a scenario would result in complete
loss of product and/or personnel protection.

Service Line Requirements

All utility service lines should be ¼ inch O.D., brass, copper, or stainless steel, and
equipped with an easily accessible shut-off valve. The service valves are rated for
operation at 40 PSI (275 kPa). If the service line pressure exceeds this, it must be
equipped with a pressure regulator to reduce the line pressure.
Note: The use of flammable gases or solvents should be avoided in the biosafety
cabinet. Open flame in the cabinet will disrupt the laminar airflow in the cabinet and may
damage the HEPA filters. Flammable gases or solvents may reach explosive
concentrations in the cabinet or ductwork. If you feel that the procedure requires the use
of an open flame or flammable materials, contact your institution's safety office.
for overall product dimensions.
Typical Operating
Current (Amps)
3 A
3 A
1.5 A
5 A
5 A
2.5 A
11
Electrical Circuit Requirements
115 V, 60 Hz, 12 A
100 V, 50/60 Hz, 12 A
230 V, 50/60 Hz, 6 A
115 V, 60 Hz, 12 A
100 V, 50/60 Hz, 12 A
230 V, 50/60 Hz, 6 A
1
1 Phase
1 Phase
1 Phase
1 Phase
1 Phase
1 Phase