GUIDANCE TABLES (continued)
Recommended separation distance between Portable and Mobile RF Communications equipment
and the Prima for ME equipment and ME systems that are not Life-supporting.
Guidance and Manufacturer's Deceleration - Electromagnetic Immunity
Recommended separation distance for between Portable and Mobile RF Communications
The Prima is intended for use in electromagnetic environments in which radiated RF disturbances are
controlled. The customer or user of the Prima can help prevent electromagnetic interference by Maintaining
a minimum distance between portable and mobile RF communications equipment (Transmitters) and the
Prima as recommended below, according to the maximum output power of the communication equipment.
Max Output Power
of Transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation Distance
(d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: At 80MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
6182001-795
equipment and the Prima
Seperation (m)
150kHz to 80 MHz
d=(3.5/v1) (Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.6666
Seperation (m) 80kHz to 800
MHz d=(3.5/E1) (Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.6666
Prima Pro
Seperation (m) 800 MHz to
2.5GHz d = (7/E1) (Sqrt P)
0.2333
0.7378
2.3333
7.3786
23.3333
Issue 1.0