PRIMA WALL MOUNT
User/Installation Manual
To ensure proper use of this instrument as well as to avoid injury while operating
Instrument, understanding this manual completely before use is highly recommended.
Part No: 6137100-795
Issue 1.3
Printed on Jan,2020

Advertisement

loading

Summary of Contents for Labomed PRIMA WALL MOUNT

  • Page 1 PRIMA WALL MOUNT User/Installation Manual To ensure proper use of this instrument as well as to avoid injury while operating Part No: 6137100-795 Instrument, understanding this manual completely before use is highly recommended. Issue 1.3 Printed on Jan,2020...
  • Page 2: Table Of Contents

    Wall mount for Prima TABLE OF CONTENT SYMBOL INFORMATION WARNINGS AND CAUTIONS INTRODUCTION DIMENSIONS OF WALL MOUNT ASSEMBLY DIMENSIONS OF WALL MOUNTED MICROSCOPE DIMENSIONS OF WALL MOUNTED MICROSCOPE WITH LONG ARM INSTALLATION OF WALL MOUNT DIMENSIONS OF WALL MOUNT 8-11 WALL MOUNT INSTALLATION INSTALLATION OF WALL MOUNT FOR WOODEN WALL INSTALLATION OF WALL MOUNT FOR CONCRETE WALL...
  • Page 3: Symbol Information

    THE SAFETY AND SATISFACTION OF USERS AND PATIENTS IS ALWAYS THE HIGHEST WARNING PRIORITY BY LABOMED. THIS MANUAL CONTAINS IMPORTANT INFORMATION REGARDING THE SAFE AND RISK FREE USE OF THIS EQUIPMENT AND SHOULD BE READ THOROUGHLY PRIOR TO INSTALL AND USE OF THE EQUIPMENT. FAILURE TO READ AND UNDERSTAND THIS MATERIAL COULD RESULT IN INJURY TO PATIENTS OR PERSONNEL OR IN DAMAGE TO THE EQUIPMENT.
  • Page 4: Warnings And Cautions

    Using these solvents also presents a danger to individuals if the solvents are opened in a poorly ventilated room. Replacement parts, such as cables, must be purchased through LABOMED to ensure CAUTION proper compliance requirements. The use of other cables may affect EMC performance.
  • Page 5: Introduction

    This Manual is only about Installation of LABOMED Wall Mounted System for Prima Series of Surgical Microscopes To Install Complete wall Mounted Prima Microscope This Manual should be used along with Microscope manual packed separately with microscope arm box.
  • Page 6 in MM 200 ±0.2...
  • Page 7: Dimensions Of Wall Mounted Microscope

    Wall mount for Prima DIMENSIONS OF WALL MOUNTED MICROSCOPE WALL MOUNT WITH STANDARD SUSPENSION ARM 196mm 525mm 600mm THRU HOLE Ø12.0mm 406mm ±150° ±100° Fig No. 1 Issue 1.3 6137100-795 Wall Mount Installation Printed on Jan,2020...
  • Page 8: Dimensions Of Wall Mounted Microscope With Long Arm

    Wall mount for Prima DIMENSIONS OF WALL MOUNTED MICROSCOPE WITH LONG ARM WALL MOUNT WITH LONG SUSPENSION ARM 525mm 922.6mm 196mm THRU HOLE Ø12.0mm 406mm ±100° ±150° Fig No. 2 Issue 1.3 6137100-795 Wall Mount Installation Printed on Jan,2020...
  • Page 9: Installation Of Wall Mount

    Wall mount for Prima INSTALLATION OF WALL MOUNT WALL MOUNT WITH STANDARD SUSPENSION ARM & SWIVEL ARM WITH 8" EXTENSION ARM Fig No. 3 Issue 1.3 Wall Mount Installation 6137100-795 Printed on Jan,2020...
  • Page 10: Dimensions Of Wall Mount

    Wall mount for Prima DIMENSIONS OF WALL MOUNT WALL MOUNT WITH STANDARD SUSPENSION ARM & SWIVEL ARM WITH 8" EXTENSION ARM 196mm 525mm 600mm THRU HOLE Ø12.0mm 406mm ±150° ±100° Fig No. 4 Issue 1.3 Wall Mount Installation 6137100-795 Printed on Jan,2020...
  • Page 11 Wall mount for Prima Wall mount: Alternative installation to the patient’s side (With standard suspension arm) 1021mm Fig No. 5 Issue 1.3 Wall Mount Installation 6137100-795 Printed on Jan,2020...
  • Page 12 Wall mount for Prima Wall mount: Alternative installation to the patient’s side (with long suspension arm) 1182mm Fig No. 6 Issue 1.3 Wall Mount Installation 6137100-795 Printed on Jan,2020...
  • Page 13 Refer to Figure 1,2,3,4,5 and 6. Measure the distance between the centers of the studs. The Labomed Wall Mount Model is designed to mount directly on a wall with wood studs spaced 406mm (16”) centers or on solid brick/concrete wall minimum 229mm (9”) thick wall.
  • Page 14 TO FIX RAVAL PLUG ASSEMBLY IN WALL AFTER FIXING RAVAL PLUG ASSEMBLY USE DRILL Ø14mm UPTO DEPTH 85mm UNSCREW BOLT WALL MOUNT BRACKET FIX WITH BOLT...
  • Page 15 Wall mount for Prima LEVELING THE WALL MOUNT ON MASONRY WALL 1. Slightly tighten the bolts on the top side- leaving the bottom bolts slightly loosened. 2. Place a level across the top of the back plate to check the horizontal leveling. See figure-13. 3.
  • Page 16 A copy of this docu- ment must be provided to labomed representative dealer. 3. The wall Post on which the Prima wall mount is to be mounted must have the following load Capacity:...
  • Page 17 Wall mount for Prima Control and Power Supply of the system Note: The maximum weight of accessories on Prima surgical microscope must not exceed 7kg. The weight of this wall mount without surgical microscope is 27Kg. Power line: 3 X 1.5mm2 Fusing: 2.5A Power consumption: max.
  • Page 18 Wall mount for Prima Confirmation of structural calculation and execution We hereby confirm that the installation of this Prima wall mount 6137100-806 from Labomed is the sole responsibility of the customers and will be in compliance with the applicable national standards and regulations.
  • Page 19 Revision History Rev. No. Date of Release ECR # Change App. By 4/4/2019 DCR/10/19 Long Arm Added S.Bal S.Bal Issue 1.3 6137100-795 Wall Mount Installation Printed on Jan,2020...
  • Page 20 920 Auburn Court Fremont, CA 94538 U.S.A. Phone: 510-445-1257 Fax: 510-991-9862 E-mail: sales@laboamerica.com www.laboamerica.com EU REP. Labomed Europe Essebaan 50 NL-2908 LK Capelle a/d IJssel The Netherlands Tel: +31 (0)10 4584222 Fax: +31 (0)10 4508251 E-mail: info@labomedeurope.com A16372 ISO13485 LABOMED...