CEILING MOUNT INSTALLATION (continued)
CONTROL AND POWER SUPPLY OF THE SYSTEM
5e
NOTE:
The maximum weight of the surgical microscope including accessories must not exceed 7kg and
the respective weight of our accessory equipment is specified in the price list.
The weight of the mount including the surgical microscope is as follow:
Ceiling mount kit
Surgical microscope
Total
Power Line
Fuse
Power Consumption
NOTE:
A socket with properly installed protective earth connection must be provided at the installation
site on Ceiling flange. The terminals for power connection are allocated on Ceiling Mount.
Potential equalization: Take the necessary actions in the building to include the instrument in the
protective measure of "potential equalization''.
NOTE:
The series 6181000 suspension systems including the components mentioned here comply safety
requirement with CE, MDD 92/42 EEC and IEC 60601-1-3.2.
Such activities as the installation of the flange and the installation of conduits and electrical lines
are the customer's responsibilities.
List of Contents: Ceiling Kit
1.
Raval Plugs
2.
Washers
3.
Ceiling Mount
4.
Column
5.
Ceiling Mount Cover
6182001-795
:
18 kg
:
20 kg
:
38 kg
: 3/1.5mm 2
: 4.0 Amp
: Max. 200W
: 06 Nos
: 06 Nos
: 01 Nos
: 01 Nos
: 01 Nos
Prima Pro