Chapter 4 Hvac; Section 1.0 Environmental Requirements; Relative Humidity And Temperature - GE Optima XR642 Pre-Installation

Table of Contents

Advertisement

GE Healthcare
Direction 5743854-1EN, Revision 1

Chapter 4 HVAC

Note: Table information is only applicable for Optima XR648.

Section 1.0 Environmental Requirements

1.1

Relative Humidity and Temperature

Table 4-1 Environmental Requirements (Relative Humidity & Temperature)
Product or Component
URP Detector
Wall Stand Std.& Man. & Ext.
Table (TBL)
OTS
System Cabinet(SKL1)
Maxiray 100-09 X-ray Tube
Radiographic Stretcher Table (optional)
Operator Console:
PC Tower
LCD Monitor
Limits for rates of change:
In-Use
< 10 degree C / hour
< 30% / hour
STORAGE values only refer to equipment that is still in shipping containers. If the equipment is partially or completely
installed, refer to IN-USE values.
The operating environment is the environmental requirements for the normal use of the device during the expected
service life. It dose not apply to the short term manufacturing testing and the like.
MIN
30%
30%
30%
30%
30%
-
30%
30%
30%
Storage
< 20 degree C / hour
< 30% / hour
Chapter 4 - HVAC
Optima XR642/XR648 Pre-Installation
RELATIVE HUMIDITY
(Non-Condensing)
IN-USE
MAX
60%
60%
60%
60%
60%
-
60%
60%
60%
Temperature
IN-USE
MIN.
MAX.
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
15° C (59° F)
35° C (95° F)
Page 93

Advertisement

Table of Contents
loading

This manual is also suitable for:

Optima xr648

Table of Contents