Fault description
No display
TACHO-ERROR 1, 2, 96
IMBALANCE 3*
CONTROL-ERROR 4.1-4.5, 6
N > MAX 5.0, 5.1
N < MIN 13
ROTORCODE 10.1-10.3
MAINS INTERRUPT 11*
VERSION-ERROR 12
CONTROL-ERROR 25.1-25.4
CRC ERROR 27, 27.1
SER I/O-ERROR 31, 34, 36
° C * -ERROR 51, 53-55
° C * -ERROR 52.0, 52.1
° C * -ERROR 58.0, 58.1
° C * -ERROR 58.6, 58.7
AB1701en_SA
Cause
No power. Triggering of the over-
current protection fuse. Triggering
of the automatic circuit breaker
(for types 1701-01 and 1706-01
only).
Tacho defective. Motor, electronics
defective.
The rotor is unevenly loaded.
Lid lock error.
Overspeed error.
Underspeed error.
Rotor coding error.
Loss of mains power during the
centrifugation run. The centrifuga-
tion run was not completed.
No conformity of the electronic
components, error/defect in elec-
tronics.
Error/defect in electronics.
Error/defect in electronics.
Error/defect in electronics.
Error/defect in electronics.
Overtemperature in the centri-
fuging chamber. Error/defect in
electronics
Temperature deviation too great.
Temperature deviation too great.
Rev.: 04 / 11.2023
Troubleshooting
Remedy
■
Check the supply voltage.
■
Switching on the automatic cir-
Chapter 9.4
cuit breaker, see ⮫
'Switching on the automatic circuit
breaker (for types 1701-01 and
1706-01 only)' on page 60 .
■
Set the mains switch to
■
Open the lid.
■
Set the mains switch to
■
Wait at least 10 seconds.
■
Turn the rotor vigorously by hand.
■
Set the mains switch to
rotor must rotate while switching
on.
■
Open the lid.
■
Check the loading of the rotor.
■
Repeat the centrifugation run.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Open the lid.
■
[START] button.
Press the
■
If required: Repeat the centrifuga-
tion run.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Increase the "Error 58 Temp"
value.
[I] .
[0] .
[I] . The
57 / 63