Fault description
FU/CCI-ERROR 60,
61.2-61.20, 61.128-61.132,
62
FU/CCI-ERROR 61.1
SENSOR-ERROR 90
SENSOR-ERROR 91-93
° C * -ERROR 97, 98
NO ROTOR OR ROTORCODE
ERROR
WRONG ROTOR !!!
N > ROTOR MAX
N > ROTOR MAX in Prog: e.g.
3
Runtime 00:00 in Prog: e.g.
B.3
Empty Program
Ramp Unit Time in Prog: e.g.
B. 3
Acc time > Run time
58 / 63
Cause
Error/defect in electronics/motor.
Mains voltage too low. Error/
defect in electronics/motor.
Error/defect in electronics.
Error/defect in imbalance sensor.
Error/defect in electronics.
No rotor installed. Tacho defective.
For type 1701-30 only: The
installed rotor is not approved for
this device.
Speed in the selected program
greater than the maximum rotor
speed.
The rotor has been changed. The
built-in rotor has a higher max-
imum speed than the previously
used rotor. The rotor has not
yet been recognised by the rotor
detection.
There is a program in the dis-
played program location where the
speed is greater than the max-
imum rotor speed.
The rotor has been changed. The
built-in rotor has a higher max-
imum speed than the previously
used rotor. The rotor has not
yet been recognised by the rotor
detection.
There is a continuous program in
the displayed program location.
There is no program link stored in
the displayed program location .
There is a program in the dis-
played program location that has
a start-up and/or run-down time.
The set start-up time is longer
than the run time.
Rev.: 04 / 11.2023
Remedy
Perform a MAINS RESET.
Check the mains voltage.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Open the lid.
Install the rotor.
Open the lid.
Install a rotor that is approved for
this device.
Check and correct the speed.
Set a speed up to the maximum
speed of the previously used rotor.
[START] button to per-
Press the
form rotor detection.
Check and correct the speed.
Set a speed up to the maximum
speed of the previously used rotor.
Press the
[START] button to per-
form rotor detection.
Replace the continuous program in
the program link with a program
with time preselection.
Open a program link.
Replace the program in the pro-
gram link with a program with a
start-up and braking stage.
Set a start-up time that is shorter
than the run time.