5.4

Switching the centrifuge on and off.

Switching the centrifuge on
Immediate display of centrifu-
gation data after switching on
Switching off the centrifuge
AB1701en_SA
NOTICE
Damage to the device due to condensation.
There is a risk of condensation forming on electrical com-
ponents when component surfaces are cold and the sur-
rounding air is warmer. The condensation that forms may
cause a short circuit and/or destroy electronics.
Warm the device up for at least 3 hours in a warm room
before connecting it to the mains.
or
Warm up for 30 minutes in a cold room.
Personnel:
Trained user
1.
A type B residual current circuit breaker must be used if the device
is additionally protected with a residual current circuit breaker in the
building installation.
When using a different type, the residual current circuit breaker may
either not switch off the unit if there is a fault on the unit, or it may
switch off the unit even though there is no fault on the unit.
2.
Check whether the mains voltage matches the specification on the
rating plate.
3.
Connect the device to a standard mains socket using the mains
cable.
Personnel:
Trained user
Set the mains switch to
�� The buttons flash, depending on the centrifuge type.
The following indicators appear one after the other, depending on
the centrifuge type:
the centrifuge model
the program version and mains voltage
the rotor code (Rotor), the maximum rotor speed (Nmax) and
the centrifuging radius (R) of the last rotor recognised by the
rotor detection.
The centrifuging radius displayed is a default value that must be
adjusted depending on the accessories used.
When the lid is closed:
When the lid is open: The centrifugation data of the last pro-
gram used or program 1.
1.
Set the mains switch to
2.
Press and hold any button at the first visual change in the display
(inverse display).
�� Centrifugation data is displayed.
The rotor is stationary.
Set the mains switch to
Rev.: 04 / 11.2023
[I] .
'OPEN OEFFNEN' indicator
[I] .
[0] .
Commissioning
25 / 63