Standby Cooling; Precooling The Rotor; Delayed Cooling - Hettich ROTINA 380 Operating Instructions Manual

Table of Contents

Advertisement

Software operation
7.4.2

Standby cooling

7.4.3

Precooling the rotor

7.4.4

Delayed cooling

40 / 63
When the rotor is at a standstill and the lid is closed, the centrifuging
chamber is cooled to the preselected temperature if this is less than 20 °C
or 68 °F.
The preselected temperature is displayed during standby cooling.
For rapid pre-cooling of the unloaded rotor and accessories, we recom-
mend a centrifugation run with the continuous run settings and a speed of
Swing-out rotor: approx. 20% of the maximum rotor speed used.
Angle rotor: approx. 40% of the maximum rotor speed used.
The centrifugation run, for precooling of the rotor, takes place automatically
with the PREC program (PRECOOLING).
A centrifugation run, for precooling the rotor, cannot be executed when
working with program links.
The rotor is stationary.
[Cooling] button.
1.
Press the
�� The button flashes until the rotor is read in for precooling.
The button lights up once the rotor has been read in.
The rotor speed or the resulting RCF value, the temperature in
the centrifuging chamber (only for centrifuges with cooling) and the
remaining or elapsed time are displayed during the centrifugation
run.
2.
Press the
[STOP/OPEN] button.
�� Precooling of the rotor is terminated.
Ramp-down takes place with the selected brake level.
The brake level is displayed.
If required, the settings can be adjusted so that cooling takes place with
a time delay after the centrifugation run has started. The delay time is
adjustable from 15 to 900 seconds, in 1 second increments. No delay time
is set ex works.
1.
Press and hold the
'***Machine Menu***' is displayed after 8 seconds.
��
[PROG] button repeatedly until '-> Settings' is displayed.
2.
Press the
[START] button.
3.
Press the
'SOUND / BELL = on' or 'SOUND / BELL = off' is displayed.
��
[PROG] button repeatedly until 'Cool acc time = 0' is dis-
4.
Press the
played.
[Rotary knob] to set the desired value.
5.
Use the
0 = no delay time
[START] button.
6.
Press the
�� The setting is stored.
'Store Settings...' is displayed briefly.
'-> Settings' is then displayed.
Rev.: 04 / 11.2023
[PROG] button.