Heating (For Centrifuges With Heating); Machine Menu; Querying System Information - Hettich ROTINA 380 Operating Instructions Manual

Table of Contents

Advertisement

Software operation
7.5

Heating (for centrifuges with heating)

enable / disable
7.6

Machine Menu

7.6.1

Querying system information

42 / 63
[Rotary knob] to set the desired value.
5.
Use the
Adjustable from 4 °C to 25 °C, in 1 °C increments, as well as the "dis-
abled" setting. Temperature monitoring is deactivated when "disabled"
is set.
[START] button.
6.
Press the
�� The setting is stored.
'Store Settings...' is displayed briefly.
'-> Settings' is then displayed.
[STOP/OPEN] button once to exit the 'Settings menu'
7.
Press the
or
Press the
[STOP/OPEN] button twice to exit the '***Machine Menu***' .
During the centrifugation run, the centrifuging chamber is heated to the
preselected temperature if required. The heating is switched off when the
rotor is at a standstill.
Swing-out rotors and angle rotors must run at maximum speed.
CAUTION
Danger of burns from hot surfaces.
The surface temperature of the heating element in the centri-
fuging chamber can be up to 500 °C or 932 °F.
Do not touch the heating element.
NOTICE
Damage to plastic buckets due to excessive temperature
Plastic buckets may only be used at temperatures up to
a maximum of 40 °C or 104 °F.
The rotor is stationary.
[T/°C] button repeatedly until 'Heater = off' or 'Heater = on'
1.
Press the
is displayed.
[Rotary knob] to set 'off' or 'on' .
2.
Use
off = heating disabled
on = heating enabled
[T/°C] button or the [START] button.
3.
Press the
�� The settings are stored.
Centrifugation data is displayed.
The following system information can be queried:
Centrifuge model
Mains voltage
Rotor information
Rev.: 04 / 11.2023
AB1701en_SA

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rotina 380 r

Table of Contents