6.2.8

Passive analog outputs

Passive analog inputs require external power supply for operation. Set analog output dip switch
to P (passive) before powering up
the system.
Exemplary usage of passive analog output
6.2.9

Active analog inputs

Active analog inputs need no external powering of the input terminal. Set analog input dip
switch to A (active) before powering up the system.
SITRANS TDL provides power supply for the active analog inputs and thus analog inputs
MUST NOT BE powered externally.
Exemplary usage of active analog input