Analog Outputs - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

8.5

[1.3] Analog outputs

Overview
Note
Availability of the menu
This menu is only available if your device has at least one of the option modules (2.1 and/ or
2.2) installed. If the option modules are not installed in the device, the "Quick Start" menu has
a corresponding menu entry.
The settings of the analog outputs are component-specific and depend on the currently set
measuring range.
You can define current ranges for valid measured values or set inverted current ranges. If you
use inverted current ranges, the assignment changes, for example: 0 ... 10 % CO ≡ 0 ... 20 mA
→ 0 ... 10 % CO ≡ 20 ... 0 mA.
To ensure that negative measured values do not influence the further processing of measured
values, negative analog output values can be suppressed. The unchanged value is output on
the device display.
In addition, you can define responses to specific device statuses. The following outputs are
available as alternatives in case of faults or with enabled function check:
● Display of the current measured value
● Display of the last measured value determined
● Display of the low or high limit of the analog output. The corresponding values are listed in
the table below in the section "Analog output range".
Quick Start
Compact Operating Instructions, 05/2015, A5E31805656-03
Commissioning: Sequence
8.5 [1.3] Analog outputs
101