Parameters Tab - Siemens SITRANS TDL Operating Instructions Manual

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

• IO
– Aout (Analog output measurement)
– Ain (Analog input measurement)
• DigitalInOut
The following sections describe how to set the measurement groups above using the
SITRANS TDL web server.
8.2.2.5

Parameters tab

The parameter tab allows configuring of external temperature or pressure signals fed in through
the analog ports, as well as configuring analog output signals from the device.
By default, there are signals assigned to the analog inputs/outputs. The user can also define the
signal in the field.
Note
For the proper measurement of the concentration, provide the actual path length determined
at the installation site (in meters) Installing/mounting (Page 25).
Edit values by clicking in the box.
Enter the desired value, and click the send button to the right of the edited field.
For drop-down menu items, once changed, the Send button is active, and clicking will save
changes in the analyzer firmware.
Parameter window
• Download parameter setting: Current parameter list can be downloaded to a *.txt file.
• Read parameters from file: Previously saved parameter list can be read from a *.txt file.
SITRANS TDL
Operating Instructions, 05/2023, A5E51429088_AA
Operating
8.2 Remote operation
105