Connecting
6.2 Connecting SITRANS TDL
6. Place two black wires in sockets
of number 25 and 26 for power‐
ing the analyzer. The same wires
should be then connected to the
24VDC power supply on the cus‐
tomer side.
7. Place other wires in proper sock‐
ets for connection with analog
and digital inputs/outputs as de‐
sired for specific application of
the analyzer. Connect with ana‐
log and digital inputs/outputs as
desired for specific application of
the analyzer.
6.2.18

Ethernet connection

A separate ethernet cable is connected via special gland for communication with the analyzer
via web server. To ensure IP65, use only the ethernet cable supplied by Siemens. Please follow
the step-by-step instruction for correct ethernet cable installation (Page 69).
68
Ethernet cable connector (emitter unit)
Ethernet cable connector (customer side)
Powering the analyzer
Connection with analog and digital inputs/outputs
Operating Instructions, 05/2023, A5E51429088_AA