Customer Cable Connection - Siemens SITRANS TDL Operating Instructions Manual

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

Connecting
6.2 Connecting SITRANS TDL
3.
Place the other two wires from the
loop cable in the ports of the termi‐
nal:
• Brown –> 1
• Green –> 2
4.
Connect the green yellow PE wire with the central grounding pin at the bottom left of the enclosure.
5.
When the installation is complete,
the assembly should look like on the
following figure.
6.2.16

Customer cable connection

Separate electrical cable is connected via power supply and I/O cable gland for powering the
analyzer and communication with the customer. Please use the cable with 0.5 mm
wires and external diameter between 7 to 12 mm (0.3 to 0.5 in) to fit into the gland. Cable
supplied by Siemens is preferred.
Please follow the step-by-step instruction for correct installation.
66
Placing wires from the loop cable in the terminal
Completed assembly
Operating Instructions, 05/2023, A5E51429088_AA
(10 AWG)
2
SITRANS TDL