List Of Termnal Blocks - Fuji Electric ZAF-6 Instruction Manual

Thermal conductivity gas analyzer thermomat
Table of Contents

Advertisement

3.5.5 Contact output (optional terminals: DO1 (11) – (12), DO2 (13) – (14), DO3 (15) – (16),
DO4 (17) – (18), DO5 (19) – (20))
Contact capacity: 250V AC/2A resistive load
• Relay contact output: Selecting ON/OFF closes/opens contact
Note)

3.5.6 List of termnal blocks

Interference compensation
Remote range changeover
Remote automatic calibration start
INZ-TN6ZAF-E
Standard terminal
Option terminal
AIN+
input 1 to 5V DC
AIN–
DI1
Remote output holding
DI2
DI2
DI3
DI3
Refer to "Input/output contact specifications" in Section 9.2 CODE SYMBOLS.
1
L
Power supply 100 to 240V AC, 50/60Hz
2
N
3
E
Grounding terminal
+
4
Measured value output (as specified)
5
4 to 20mA DC
0 to 1V DC
(M4 screw)
0 to 10mV DC
1 11
DO1
Contact output 1
2 12
DO1
DO2
3 13
Contact output 2
DO2
4 14
DO3
5 15
Contact output 3
DO3
6 16
DO4
7 17
Contact output 4
DO4
8 18
DO5
9 19
Contact output 5
DO5
10 20
(M3.5 screw)
.