6.7.2 Span calibration
more of the range value.
(1) Press the
MODE
screen.
(2) Pressing, the
for span calibration.
Caution
If "Both" has been selected for "Calibration
range action", calibration is carried out in both
ranges simultaneously.
(3) Wait until the indication is stabilized
is supplied. After the indication has
been stabilized, press the
calibration of range selected by the cursor
is performed.
To close "Span Calibration"
To close the "Span Calibration" or
cancel this mode midway, press the
key. A previous screen will return.
INZ-TN6ZAF-E
key on the measurement
ENT
SPAN
 
ESC
2013-02-15 17:24
H
1.28
2
0
2
4
2.00
SPAN Cal.
ENT : Gas flows
ESC : Back
Range1
0–3 vol%
H
2
Range2
0–10 vol%
SPAN Cal.
ENT : Go on calibration
ESC : Not calibration
Range1
0–3 vol%
H
2
Range2
0–10 vol%
Low alarm
vol%
6
8
10 vol%
8.00
1.28
1.28
45