appears, select "Both" or "Current" (for 2
ranges) and press the
• When selecting "Current", the range
alone displayed is calibrated.
To close "Setting of Calibration Range"
To close "Setting of Calibration Range" or
to cancel this mode midway, press the
A previous screen will return.
Note) When calibration is performed by the "Both" setting under the normal operating condition,
prepare a span gas cylinder on the normal operating range side. It is recommend to
perform span gas calibration in the normal operating range.
24
key.
ENT
ESC
In case of "both" setting
ZERO Cal.
Range1
H
2
Range2
Cal. Settings
Cal. Range
Range1
H
2
Range2
key.
by the
or the
ENT : Gas floes
ESC : Back
0–3 vol%
1.28
0–10 vol%
Set calibration range
current or both range
0–3 vol%
c u r r e n t
0–10 vol%
key.
-
pear in both ranges
INZ-TN6ZAF-E