Contact operation selection items and alarm display
Contact operation
selection item
(1) Upper limit value
(2) Lower limit value
(3) Upper or lower
limit value
(4) Upper limit value
+ Lower limit value
(5) Two-step upper
limit value
(6) Two-step lower
limit value
message is displayed above the measured value.
•
28
Bar graph and alarm display screen
0
20
40
60
Alarm range display (1) to (6)
Upper limit 1 alarm
Lower limit 1 alarm
Upper or lower
limit alarm
Upper limit 1 alarm
Lower limit 1 alarm
Upper limit 1 alarm
Upper limit 2 alarm
Lower limit 1 alarm
Lower limit 2 alarm
80
100
* Refer to "6.6.5 Contact output
setting" for the setting and
assignment of contact outputs.
One point
(Select one upper limit 1 alarm.)
One point
(Select one lower limit 1 alarm.)
One point
(Select one upper limit or
lower limit alarm.)
Two points
(Select one upper limit 1 and
one lower limit 1 alarms.)
Two points
(Select one upper limit 1 and
one upper limit 2 alarms.)
Two points
(Select one lower limit 1 and
one lower limit 2 alarms.)
2013-02-15 17:24
H
1.28
2
0
2
4
6
2.00
Contact output
Low alarm
vol%
8
10 vol%
8.00
INZ-TN6ZAF-E