User's Manual
CAT 0
CAT I
CAT II
CAT III
CAT IV
Certification
Certification
Warning
The UTE310 Digital Power Meter only supports measurement of power supplies under CAT II (600V)
type overvoltage conditions, so please use the instrument strictly in accordance with this
measurement environment.
General Instructions
The following instructions are listed for personal safety and equipment protection.
Defective protection function
Instruments.uni-trend.com
casing cover
Grounding
Frame and
Grounding
Protective grounding terminal
Grounding
Measurement grounding terminal
This instrument is suitable for measurements on circuits that are not directly connected
to the grid power supply and circuits that are specially protected to be powered from the
(internal) grid. In the latter case, the transient stresses are different, and the instrument
should be used for this type of measurement to ensure that the peak transient voltage is
less than 3000 V.
Secondary electrical circuit connected to wall sockets through transformers or similar
equipment, such as electronic instruments and electronic equipment; electronic
equipment with protective measures, and any high-voltage and low-voltage circuits,
such as the copier in the office.
Primary electrical circuit of the electrical equipment connected to the indoor socket via
the power cord, such as mobile tools, home appliances, etc. Household appliances,
portable tools (e.g. electric drill), household sockets, sockets more than 10 meters away
from CAT III circuit or sockets more than 20 meters away from CAT IV circuit.
Primary circuit of large equipment directly connected to the distribution board and
circuit between the distribution board and the socket (three-phase distributor circuit
includes a single commercial lighting circuit). Fixed equipment, such as multi-phase
motor and multi-phase fuse box; lighting equipment and lines inside large buildings;
machine tools and power distribution boards at industrial sites (workshops).
Three-phase public power unit and outdoor power supply line equipment. Equipment
designed to "initial connection", such as power distribution system of power station,
power instrument, front-end overload protection, and any outdoor transmission line.
CE indicates a registered trademark of EU
UKCA indicates a registered trademark of UK
This product complies with the marking requirements of WEEE Directive (2002/96/EC).
Waste
This additional label indicates that this electrical / electronic product must not be
discarded in household waste.
This environment-friendly use period (EFUP) mark indicates that dangerous or toxic
substances will not leak or cause damage within this indicated time period. The
EFUP
environment-friendly use period of this product is 40 years, during which it can be used
safely. Upon expiration of this period, it should enter the recycling system.
grounding terminal
UTE310 Digital Power Meter