Chapter 5 Measurement Settings

5.1 Range Settings

5.1.1 Voltage Range Settings

·Steps
1. In any interface, press the [Voltage] key to pop out the selection window of voltage range;
2. Press the [▲] or [▼] key to select the voltage range;
3. Press the [Enter] key to save the selected voltage range and exit the selection window or wait for
10s to automatically save and then exit.
·Explanations
Selection range: Auto, 15V, 30V, 60V, 150V, 300V, 600V (CF=3)
Auto represents the automatic range.
When CF=6 or 6A, all ranges will be reduced to half of the original, i.e. 7.5V, 15V, 30V, 75V, 150V, 300V.

5.1.2 Current Range Settings

·Steps
1. In any interface, press the [Current] key to pop out the selection window of current range;
2. Press the [▲] or [▼] key to select the current range;
3. Press the [Enter] key to save the selected current range and exit the selection window or wait for
10s to automatically save and then exit.
Common Measurement
Selection range: Auto, 5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 500mA, 1A, 2A, 5A, 10A, 20A.
The display interface will synchronize the currently selected current range.
Auto represents the automatic range.
When CF=6 or 6A, all ranges will be reduced to half of the original, i.e. 2.5mA, 5mA, 10mA, 25mA, 50mA,
100mA, 250mA, 500mA, 1A, 2.5A, 5A, 10A.
When use EXT CH (the external current sensor) to measure
When using Ext1, the range can select to Auto, 2.5V, 5V, 10V
When using Ext2, the range can select to Auto, 50mV, 100mV, 200mV, 500mV, 1V, 2V.