User's Manual
F2, F3 corresponding to the function indication. For the sake of simplicity, respectively, use F1, F2, and F3 to
describe its function in this subsection.
Explanation
F4: Press the "F4" key to display SEL-A, SEL-B, SEL-C, SEL-D in sequence, which means the display area A,
B, C, D is selected, and then press "F5" to edit the selected area, and the border of the window will change
from white to red when the window is selected;
F5: FUNC, switch the displayed parameter, when math is selected in the C area, the parameter displayed in
the C area is the result of the operation of the A area and the B area, and the mathematical operation
formula is determined by the mathematical formula in "F6";
F6: Press the "F6 key to display CF_U (voltage crest factor), CF_I (current crest factor), A+B, A-B, AxB, A/B,
A/BB, AA/B, AV_P (average active power during integration, the specified average active power will only
be displayed when the integration in the start/stop state) in sequence.
Example of mathematical operation
AxB: Display the result of multiplying area A with area B
A/B: Display the result of dividing area A with area B
Calculating the absolute value of impedance (Z)
Display Area A
A/BB: Display the result of dividing the square of area A with area B
Instruments.uni-trend.com
This function can be used to display apparent power S in the C area in VIEW-3 style.
Display Area A
Display Area B
U
Display Area B
U
I
Calculating the impedance (Z), the resistance (R) and the reactance (X)
Display Area A
Display Area B
Display Area C
I
U*I=S
Display Area C
U
= | Z |
I
Display Area C
UTE310 Digital Power Meter
Wiring Method
No limit
Wiring Method
Wiring Method
42 / 99