3.0 Power ON Test

Equipment Needed
Power Supply 12V 10A,
AED PLUS, Quick-Test, Test Fixture (P/N 9100-0567-TF),
Battery Current Test Fixture (P/N 9100-0690-TF) or AED Pro Disposable Battery (P/N
1008-002003-01).
Setup
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Test
3.1
3.2
9650-0309-01 Rev. E
Action
Place the AED Pro Battery in
the battery well (then turn on the
Power Supply if Battery Current
Test Fixture is used)
Press and hold the On/Off
button for at least 7 seconds,
then release it.
ZOLL AED Pro Service Manual
Expected Result(s)
Initial Power up. Unit will turn off
by itself after UNIT OK prompt.
Within 10 seconds after the unit
is turned on, the Ready For Us
indicator displays a green
Check. Indicating the unit is
ready for use. A red "X"
indicates the unit is not ready for
use.
Power ON Test
Pass
Fail
1–5