Table 2-1. General Issues (continued)
Symptom
RELEASE SHOCK BUTTON
PLUG IN DEFIB CABLE
RELEASE LEFT SOFTKEY
or
RELEASE RIGHT SOFTKEY
9650-0309-01 Rev. E
Operator Action
— Release the Shock button. Wait
until the unit issues the prompt
PRESS FLASHING SHOCK BUTTON
before pressing the button.
Note: This prompt appears if an ECG
cable is connected, but the unit
is not configured for ECG
monitoring.
— Check the cable and replace if
necessary.
— Release the softkey.
ZOLL AED Pro Service Manual
Troubleshooting the AED Pro Unit
Technical Action
None.
None.
None.