Power On Self Test

The ZOLL AED Plus unit performs a self test whenever the unit is turned on and verifies the following
functions:
1. Battery Capacity: Verifies that the battery usage indicator shows at least 50% of full battery
capacity remaining.
2. Defibrillation Electrodes Connection: Verifies that the defibrillation electrodes are properly pre-
connected to the device.
3. ECG Circuitry: Verifies that the ECG signal acquisition and processing electronics are functional.
4. Defibrillator Charge and Discharge Circuitry: Verifies that the device's defibrillator electronics
are functional and can charge and discharge at 2 joules.
5. Microprocessor Hardware/Software: Verifies proper function of the ZOLL AED Plus
microprocessor electronics and the integrity of its software.
6. CPR Circuitry and Sensor: Verifies that CPR monitoring and compression depth detection are
functional.
7. Audio Circuitry: Verifies that voice prompts are functional.

Manual Self Test

You can initiate a manual self test on the ZOLL AED Plus unit by pressing and holding the unit's On/
Off button for 5 seconds. The ZOLL AED Plus illuminates all graphic indicators and issues voice and
LCD display messages to allow user verification of the device's visual and auditory output
functionality. In addition, the LCD shows the application software version currently running on the
unit.
Boot File Version
Language File Version
This self test verifies the following ZOLL AED Plus functions:
1. Battery Capacity: Verifies that the battery usage indicator shows at least 50% of full battery
capacity remaining.
2. Defibrillation Electrodes Connection: Verifies that the defibrillation electrodes are properly pre-
connected to the device.
3. ECG Circuitry: Verifies that the ECG signal acquisition and processing electronics are functional.
4. Defibrillator Charge and Discharge Circuitry: Verifies that the device's defibrillator electronics
are functional and can charge and discharge at 2 joules.
5. Microprocessor Hardware/Software: Verifies proper function of the ZOLL AED Plus
microprocessor electronics and the integrity of its software.
6. CPR Circuitry and Sensor: Verifies that CPR monitoring and compression depth detection are
functional.
7. Audio Circuitry: Verifies that voice prompts are functional.
8. Display: Verifies that visual indicators are functional.